OTF-1200X-S is a split furnace designed for easy replacement of heating module and heating element for any application up to 1100˚C. Here is the short summary for procedure of heating module replacement
Disassemble of Heating Module:
- Remove the stainless-steel plates on the upper and lower chamber
- Remove both upper side-panels;
- Remove the thermocouple and disconnect the power cable from the upper left panel side
- Remove both lower side-panels
- Disconnect the power cable from the lower left panel side
Crimp the wire terminals before installation
Installation of Heating Module:
- Place lower heating module in bottom chamber
- Connect and secure the crimped power cable on lower left side
- Install both lower side-panels
- Install the lower stainless-steel plates
- Place the furnace side down
- Place the upper heating module in the upper chamber
- Drill a hole for thermocouple in the left side of heating module
- Install the thermocouple and connect to its power cable
- Install both upper side-panels. Please notice there is a notch on the upper left panel for thermocouple
- Install the upper stainless-steel plates
Please refer to the video for further details