 KSL-1700X-MGI-8-AD consists of eight box furnaces together for high throughput heat treat processing of new material discovery.  The temperature of each heating chamber is controlled independently up to 1700℃.   The door of each furnace is automatically opened and closed as programmed via a touch-screen computer.  Combined with a robot arm,  the furnace system can work 24/7 autonomously.

SPECIFICATIONS:

Furnace Structure

  • Integrates eight  box furnaces  ( 120x120x120 mm ) for space and energy saving 
  • Each heating chamber is independently temperature controlled, with a maximum temperature of 1700℃
  • Can simultaneously accommodate four samples for heat treatment at different temperatures
  • The door of the heating chamber can open and close automatically as programmed for sample loading and unloading
  • An optional robot arm is available for 24/7 autonomous operation
Specifications
  • Voltage: AC220V+/- 10%   50/60HZ,  Single-phase 220V
  • Power: 20 KW
  • Heating rate: ≤10/min
  • Chamber size: 120*120*120mm (  capacity 1.7 L )
  • Effective uniform heating space:  80x80x80 mm within +/- 5oC
  • Heating element: MoSi2
  • Thermocouples: B-type 
  • 8 gas inlets are installed at the back of the furnace

Working Temp. 

  • Continuous Working Temperature: 1600°C
  • Maximum Working Temperature 1700°C (< 30 mins)
  • Maximum 1750°C upon request at extra cost

Control of Temperature  and door 


  • Touchscreen temperature and door control
  • Eight independent temperature control channels enable simultaneous one-touch control
  • PID automatic temperature control system
  • Intelligent 50-stage programmable control
  • The door of each heating chamber can be opened or closed as programmed via the touch screen panel
  • Default DB9 PC communication connection port for robot operation
  • Temperature control accuracy: ±1. (optional European temperature control accuracy up to ±0.1)

Automation  option

  • A robot arm and sample tray station may be connected with the 8 chamber furnace to achieve operation autonomously.
  • Call MTI for your automation application
Compliance
  • CE certificated
  • NRTL certification is available at an extra cost

Warning &
Application note

  • MTI does not recommend customers use flammable, explosive, or toxic gases(materials). MTI is not responsible for issues arising from the use of flammable, explosive, or toxic gases(materials)
  • The equipment should operate on an independent power source, with a voltage range requirement within ±8%. It should not share the power grid with equipment generating high magnetic fields like medium- and high-frequency devices. It should not be in the same room as such equipment. 
  • Keep the equipment away from areas prone to liquid splashing
  • Keep the equipment away from areas with magnetic or conductive powders
  • Cracks in insulation material during sintering are normal; they can be repaired using patching agents. Heating elements are also consumables. If the electric furnace cannot reach the working temperature or shows a significant slowdown in heating after prolonged use, contact the manufacturer to determine if replacement is necessary
  • There might be temperature fluctuations during low-temperature(below 300) operation
  • DO NOT open the furnace at temperatures exceeding 500

 

MTI offers a One-Year limited warranty for this product.

The MTI One Year Limited Warranty does NOT cover

  • Consumable parts, such as graphite dies, O-rings, crucibles, screws, washers, glassware, etc.
  • Any damage and rust resulting from improper storage or maintenance.
  • Any damages caused by the use of corrosive and acidic gases.
  • Any damage resulting from failing to adhere to the individual product's specific operation requirements, such as water cooling, venting, etc.
Manufacturer Part Number: