Furnace Construction |
- Double-layer steel casing with air cooling
- High-purity alumina fiber insulation for max. energy saving
- Splittable cover for easy tube replacement
|
Power Consumption
|
5 KW
|
Input Voltage
|
AC 208-240V Single Phase, 50 or 60 Hz
|
Maximum Temperature
|
1200oC
|
ContinuousTemperature
|
1100oC
|
Maximum Heating Rate
|
<= 20 oC /min
|
Heating Zone Length
|
- Three zones: (880mm, 35" in total)
- Zone 1: 220mm (8.6" )
- Zone 2: 440mm (17.3")
- Zone 3: 220 mm (8.6")
|
Constant Temperature Zone
|
- If three zones are set at the same temperature, the constant temperature zone is 25" (625 mm) long with a temperature discrepancy of +/- 1oC.
- If using only one zone (center zone), the constant temperature zone is 4.3" (110 mm) long with a temperature discrepancy of +/- 1oC.
|
Temperature Accuracy
|
+/- 1oC
|
Heating Elements
|
Fe-Cr-Al Alloy doped by Mo
|
Temperature Controller
|
- Three precision temperature controllers to control three zones separately.
- PID automatic control with 30 programmable segments for precise control of heating rate, cooling rate, and dwell time.
- Built-in overheating & broken thermocouple protection.
- Over temperature protection and alarm allows for operation without attendant(s).
- +/- 1 ºC temperature accuracy.
- RS485 Communications Port.
- Optional:
- The furnace may be operated by PC by installing three control modules. Please order the MTS-02 control modules at an extra cost.
- The new version of PC module is now compatible with Lab View 05/08/2014
- If you order the PC control module together with the furnace, we will install and test it before shipping. We recommend you to order a computer laptop from us to avoid software installation problems. For more information please click here.
|
Quartz Tube& Sealing Flanges
|
-
- Vacuum flanges with valves and pressure gauges are included.
- The thermal block must be placed inside the tube as in the drawing below
|
Vacuum Sensor & Display
|
|
Sliding Flange with Sample Holder
|
|
Vacuum Sealing Flanges
|
- Two stainless steel flanges with both vacuum gauge and valves included.
- Min. vacuum pressure achieved by the mechanical pump is 5.5 m-torr. ( 5.5x10^-3 torr). The pressure can reach 10^-5 torr by using a mechanical + molecular pump.
- The leaking rate is < 5 m-torr / min. and stable vacuum pressure after 24 hours is < 2 torr.
- Click the picture below to get a details chart and data measured by CE certified vacuum meter:
(For a whole 24 hours curve display, the time segment through 08:00:00~21:10:00 in X axis is passed over in the chart, therefore, steep variety in there)
|
Oxygen Monitoring
|
An Oxygen Sensor can be used to monitor the oxygen level of gas(es) used in CVD systems for preventing or reducing oxidation. Please click the picture below to learn more:
|
Vacuum Pump
|
- 120 L/m Rotary Vane Vacuum Pump is included with max. vacuum pressure of 10^-2torr. Stainless steel hose tube is included for connecting to the tube furnace.
- Option: For precise control of the vacuum pressure, we recommend you choose our turn-key solution VPC-MV. This is a vacuum pressure regulating system with a motorized valve and gauge. It controls pressure range from 10^-6 torr ~ 760 torr.
|
Gas Flow Meter and Pressure Meter
|
-
Three gas flow meters (4% accuracy FS) are installed inside the mobile case to enable monitoring of gas flowing rates (three different kinds of gases in total).
- Compact direct read flow meter, 10-100 cc/min.
- Compact direct read flow meter, 16-160 cc/min.
- Compact direct read flow meter, 25-250 cc/min.
-
Four pressure meters are installed inside the mobile cart which is able to monitor three types of flowing gas pressure as well as the pressure of the mixing tank.
|
Valves |
Four stainless steel valves are installed in the side panel of the mobile cart to control the mixing of three inlet gases separately |
Laptop, software & WiFi Control (Optional)

|
- Brand new laptop with Microsoft Windows 10 and Microsoft Office 2013 (30 days free trial) for immediate use.
- Labview-Based Temperature Control System (MTS01) enables users to edit temperature profiles, manage heat-treatment recipes, and record and plot data for MTI furnaces.
|
Compliance |
- CE Certified
- UL / MET / CSA Certification is available upon request, additional charges will apply
|
Application Notes
Click here to learn about the installation of a gas regulator
|
|