OTF-1200X-S-VT is a compact-split vertical tube furnace with vacuum-sealed 1", 2" or 2.36" (60mm) diameter quartz tube. It provides fast heating and quenching as well as the capability to heat samples under vacuum or flowing gas atmospheres. A precision temperature controller can provide up to 30 segments of the heating, cooling, and dwelling steps with +/- 1ºC accuracy. It is an ideal furnace for carrying out quenching tests.
SPECIFICATIONS
Furnace Casing
|
|
Power
|
1.2 KW; 110VAC, 10 A or 208-240VAC, 6A, 50/60Hz
|
Max. Temperature
|
1100ºC
|
Continuous Temperature
|
1000ºC
|
Max. Heating Rate
|
<= 20ºC/min
|
Tube Size and Materials
|
|
Heating Zone Length
|
200mm (single zone)
|
Constant Heating Zone
|
60mm (+/-1ºC)
|
Temperature Controller
|
- 30 programmable segments for precise control of heating rate, cooling rate and dwell time.
- Built-in PID Auto-Tune function with overheating & broken thermocouple protection.
- Over-temperature protection and alarm allow for operation without attendant(s).
- +/- 1ºC temperature accuracy.
- You may upgrade the controller to Eurotherm 3000 with +/- 0.1ºC accuracy
|
Heating Elements |
Fe-Cr-Al Alloy doped by Mo
|
VacuumSealing
|
|
Fitting Ports
|
- Standard 1/4‘’ Barbed fittings with needle valves.
- For supplying high-pressure gas, you may replace the standard barbed hose fitting with the 304SS 1/4" Swagelok® Tube Fitting, please click underlined to order.
- Optional: In order to increase the throughput and achieve high vacuum, you may replace the standard barbed hose fitting with the KF25 adapter, please click the underlined to order.
- Optional: You may also consider using the parts below to increase the vacuum efficiency
|
Vacuum Level
|
10-2 Torr by a mechanical pump
|
Laptop, software & WiFi Control (Optional)
 
|
- Brand new laptop with Microsoft Windows 11 and Microsoft Office 2013 (30 days free trial) fo
-
Labview Based Temperature Control System (MTS01)enables users to edit temperature profiles, manage heat-treatment recipes, and record, and plot data for MTI furnaces.
- A wireless remote control provides up to 50 meters of operating range.
|
Compliance
|
- CE certified
- NRTL or CSA certification is available upon request at the extra cost
|
Warning
 Click here to learn the installation of a gas regulator
|
- Tube furnaces with quartz tube are designed for using under vacuum and low pressure < 3 psig.
- Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSIG for safe operation. Click here to learn the installation of a gas regulator.
- Vacuum pressures may only be safely used up to 1000°C
- The flow rate for gasses should be limited to<200 SCCM for reducing thermal shocks to the tube
|