MIST-1600-6 is a MIST/LPCVD system designed for the epitaxial growth of thin films on various substrates up to 1600°C heated by induction. It features an induction-heated graphite substrate holder with precise PID control. The boron nitride (BN) shower head enables a uniform deposition of precursors on a substrate at an elevated temperature. The system also comes with a hermetic quartz chamber for LPCVD without worries of precursor leaks. It is a unique tool for high-temperature thin film growth, especially for next-generation semiconductors, such as Ga2O3 and BN.
SPECIFICATIONS:
|
Power
|
- 208-240 VAC, 50/60 Hz, 3 phase
- 15 kW
- A 10-foot power cord is included. The power plug is not covered
|
|
CVD Chamber
![]()
![]()
|
- Quartz chamber: Ø8.5" OD × 20" H (Ø216 mm OD × 500 mm H)
- 6" graphite wafer stage: Ø165 OD × Ø155 ID × 50 mm H with an Ø8 mm hole at the back.
- Thermal insulation: graphite felt
- Gas lines: 1/4" compression tube fittings as gas inlet and outlet.
1/4" compression tube fittings as precursor inlet. KF16 port for Pirani vacuum gauge. pressure relief valve @ 3 psi.
- Thermocouple: 1/4" thermocouple feedthrough with a C-type TC.
- Observation window: Ø35 mm
- Vacuum port: KF25
|
| Molecular Doser (Shower Head) |
- 4" boron nitride shower head (orifice: Ø0.5 mm) designed for high-temperature uniform coating.
-
|
|
Precursor Evaporator (Bubbler)
![]()
|
|
|
Induction Heater
![]()
|
- 15 kW 3 phase 208 -240 VAC induction power supply.
- 30-100 kHz.
- Built-in interlocks of water flow, overheating, and overpowering.
- Coil: Ø220 mm ID × 70 mm H, 8 mm copper, 4 turns.
- Cooling water requirement: >30 psi, >6 L/min, T < 30°C
|
|
Temperature Control
|
- Working temperature 1500°C; 1600°C max (<1hr)
- 30 segments for PID programmable heating and cooling
- Built-in overheating protection and thermocouple failure protection.
- Temperature accuracy: ±1°C
- Thermocouple: C type, 400 mm L.
|
|
Pressure Gauge
|
|
|
Vacuum Generation (Optional)
![]()
|
-
Pfeiffer oil-free dry scroll pump, 200 L/min
- 110 VAC or 220 VAC
- A cold trap is recommended to condense the volatile precursors.
|
|
Water Chiller (Optional)
|
-
Digital Temperature Controlled Recirculating Water Chiller 58L/min, 17K BTU/hr - EQ-KJ6200
|
|
Gas Mixer (Optional)
|
|
|
Gas Detector (Optional)
|
|
Dimensions
|
|
|
Net Weight
|
|
|
Warranty
|
- Limited 1 year. Consumables are not included in the warranty.
|
|
Application Note
|
- MTI recommends placing the equipment inside a fume hood in case of any precursor leaks.
|
Operation Video
![]()
|
![]() ![]()
|