1700°C Hydrogen Gas Tube Furnace with a 60mm Alumina Tube and Hydrogen Detector - GSL-1700X-S60HG

1700°C Hydrogen Gas Tube Furnace with a 60mm Alumina Tube and Hydrogen Detector - GSL-1700X-S60HG

Minimum Order: $20 (equipment), $100 (crystals)

Shipping and handling will be applied

Price: RFQ

SKU: 

Choose Your Variant or Type in Any Add-Ons:



View full details

GSL-1700X-S60HG is a high-temperature tube furnace designed for processing materials under a hydrogen atmosphere at temperatures up to 1700°C. It features a 60 mm outer-diameter alumina processing tube, a hydrogen gas detection system, and an automatic gas inlet shutoff valve.

The furnace can also be used with inert gases and oxygen. Its precision temperature controller provides 30 programmable segments for controlling heating rates, cooling rates, and dwell times.

A hydrogen detector is integrated with the gas inlet solenoid valve to automatically shut off the hydrogen supply when the configured gas concentration alarm threshold is reached. The furnace must be operated under an appropriate fume hood with the required gas-handling safety controls.

SPECIFICATIONS

Furnace Structure
  • Double-layer steel casing with air cooling to maintain the furnace surface temperature below 60°C
  • High-purity fibrous alumina insulation for energy efficiency
  • Integrated hydrogen detector and gas inlet solenoid valve
  • Must be installed under an appropriate fume hood

GSL-1700X-S60HG hydrogen atmosphere tube furnace

View MTI EQ-FH-36 fume hood specifications.

Power
  • Maximum power: 3 kW
  • Required circuit breaker: 20A
Voltage AC 208–240V, single phase, 50/60 Hz
Maximum Temperature 1700°C (less than 1 hour)
Continuous Operating Temperature 1600°C
Maximum Heating Rate
  • Maximum: ≤10°C/min
  • Recommended: 5°C/min to reduce thermal shock to the alumina tube
Processing Tube
  • High-purity alumina (Al2O3) ceramic tube with purity greater than 95%
  • Inner diameter: 51 mm
  • Outer diameter: 60 mm
  • Length: 800 mm
  • Two fibrous ceramic tube blocks included to reduce heat radiation from inside the tube
  • Ceramic blocks must be fully inserted into the furnace chamber before heating.
  • Optional quartz tube blocks are available for applications requiring a cleaner processing environment and improved vacuum performance.
  • For 60 mm tubes, the manufacturer recommends purchasing two pairs of tube supports to help prevent tube bending at high temperatures.

Fibrous ceramic insulation tube block Ceramic tube block installation example Optional quartz tube insulation block Alumina tube support arrangement

View alumina insulation blocks.

Heating Zone Length 125 mm (5")
Constant Temperature Zone 50 mm (2") with ±2°C temperature uniformity
Temperature Controller
  • MET-certified programmable temperature controller
  • PID automatic control using SCR (Silicon Controlled Rectifier) power control with phase-angle firing and a current-limiting resistor
  • 30 programmable segments for precise control of heating rates, cooling rates, and dwell times
  • Built-in PID auto-tune function
  • Over-temperature alarm and protection
  • Broken thermocouple protection
  • Temperature accuracy: ±1°C
  • RS-485 communication port

Optional Advanced Temperature Control:

Temperature Accuracy ±1°C
Vacuum Flanges & Gas Inlet
  • One pair of stainless steel vacuum flanges installed at both ends of the processing tube
  • Manufacturer-stated vacuum capability: up to 10-5 torr
  • Manufacturer-stated flange positive-pressure capability: 5 PSI
  • Solenoid valve installed on the left flange for hydrogen gas inlet control
  • Solenoid valve connected to the hydrogen detector for automatic inlet shutoff when a leak alarm is triggered
  • Pressure gauge installed at the gas inlet
  • Stainless steel needle valve for adjusting inlet gas pressure

Important: The general operating safety instructions limit positive pressure to less than 3 PSI. The 5 PSI flange specification must not be treated as the permitted operating pressure.

Hydrogen furnace left-side gas inlet flange Hydrogen furnace right-side outlet flange Hydrogen inlet solenoid shutoff valve

Gas Outlet & Flowmeter
  • Float-type flowmeter installed at the gas outlet
  • Flowmeter range: 16–160 mL/min
  • 1/4" diameter stainless steel outlet tube provided for the manufacturer's hydrogen exhaust combustion arrangement
  • Long lighter included in the standard package for manual ignition of hydrogen at the designated outlet
  • 6" long-neck flexible safety lighter listed as included

Safety: Hydrogen exhaust ignition and disposal must follow the manufacturer's validated operating procedure and the laboratory's approved hydrogen safety requirements.

Hydrogen furnace gas outlet and flowmeter assembly

Hydrogen Detection System (Included)
  • Honeywell UL-approved Sensepoint hydrogen gas detector integrated into the system
  • Connected to the gas inlet solenoid valve for automatic hydrogen supply shutoff
  • Specified detector temperature range: -40°C to +65°C (-40°F to +149°F)
  • Hydrogen alarm threshold: 20% of the lower explosive limit (LEL)
  • When the detector reaches the configured alarm threshold, the hydrogen inlet valve automatically closes.

Honeywell hydrogen gas detection system

Heating Elements Super 1800-grade molybdenum disilicide (MoSi2) heating elements, 30 × 270 mm
Laptop, Software & Wireless Control (Optional)
  • Laptop configuration described with Microsoft Windows 10 and Microsoft Office 2013 (30-day trial)
  • EQ-MTS01 LabVIEW-based temperature control system for editing temperature profiles, managing heat-treatment recipes, and recording and plotting furnace data
  • Optional wireless remote control with a stated operating range of up to 300 meters

Optional wireless furnace control system Optional computer temperature control software

Compliance
  • CE certified
  • NRTL or CSA certification available upon request at an additional cost
Safety Warnings & Operating Requirements
  • Alumina tube furnaces are designed for vacuum and low-pressure operation below 0.02 MPa (0.2 bar / approximately 3 PSI).
  • A two-stage pressure regulator must be installed on the gas cylinder to limit the gas supply pressure to below 3 PSI.
  • Never heat the furnace while the gas release valve is closed and positive pressure exists inside the processing chamber.
  • Monitor the chamber pressure gauge throughout heating. The manufacturer's instructions require immediate pressure relief if chamber pressure exceeds 3 PSI.
  • Gas flow rates must remain below 200 SCCM (200 mL/min).
  • The furnace must be operated under an appropriate fume hood.
  • MTI requires first-time buyers to complete safe-operation training at its facility, provided free of charge.

Two-stage gas cylinder pressure regulator

View gas regulator installation information.

View MTI EQ-FH-36 fume hood specifications.

Furnace Dimension:

  • 445 x 380 x 615 mm (17.5" x 15.0" x 24.5")

Net Weight:

  • 230 lbs

Shipping:

  • Shipping Dimension: 48"x 40"x 45"
  • Shipping Weight: 280 lbs

MTI offers a One-Year limited warranty for this product.

The MTI One Year Limited Warranty does NOT cover

  • Consumable parts, such as graphite dies, O-rings, crucibles, screws, washers, glassware, etc.
  • Any damage and rust resulting from improper storage or maintenance.
  • Any damages caused by the use of corrosive and acidic gases.
  • Any damage resulting from failing to adhere to the individual product's specific operation requirements, such as water cooling, venting, etc.