Features:
  • Super Accurate, Super Stable, Super Value
  • For Use with Temperatures Up to 1335 oC (2440 oF) FOR OTF-1200X-RTP-II
  • Excellent Long Term Stability at High Temperature
  • Nickel-Chrome Based Super-alloy Sheathing Provides Excellent High Temperature Performance
  • Detective probe make temperature measure much more precise measurement.
  • Excellent Oxidation, Carburization and Chlorination Resistance
  • Made with materials of Special Limits of Error. Perfect for calibrating MTI tube furnaces using a precision temperature calibrator (matched with male plug; related product below )

    Specifications

    • Length: 18"
    • Diameter of sheath: 1/4" diameter
    • Positive Material: Ni-Cr
    • Negative: Ni-Al
      Highest Temperature: 1200 oC
      Material: Ni-Al
    • Lowest Temperature: -200 oC
    • 1/4" thermocouple can be inserted into processing tube via flange and calibrating tube furnace temperature under vacuum *(click picture to order the flange)
    Application Notes
    • If necessary, please check out K type thermocouple extension wire with mini male plug. Using compensating wires is equivalent to stretching them to keep them operating at a constant temperature. The advantages of using compensating cables are saving material costs and solving the problem of constant temperature, etc.
    • There are 2 polarities for K type thermocouple: positive and negative. Please check polarity of K type thermocouple before use. Wrong polarity may cause temperature measuring error.
    • It is also easy to test the polarity of K type thermocouples. The negative wire is MORE magnetic than the positive wire. Just put a magnet up to each wire. One will be more magnetic than the other.
    Warning
    • Chromel/Alumel is vulnerable to sulfidic corrosion which can result in a thermocouple failure. Apply a thin layer of Zirconia coating will protect the TC from corrosion and thus extend its service life. Re-apply the coating if there is a sign of layer flaking/peeling after the process. Please click the picture in below to order the product.

    1800C (3270F) Hi-Purity Zirconia Coating - EQ-634-ZO-LD