OTF-1200X-II-H2 is a compact hydrogen tube furnace for the heat treatment of oxidation-sensitive materials. The two heating zones provides a temperature gradient or a wide range of temperature uniformity. Flange fixtures and quartz tube covers are installed to provide additional safety protection in case of emergency. A Honeywell hydrogen leak detector is applied to trigger the close of the hydrogen inlet valve and the power cutoff of heating elements immediately once hydrogen leak is detected.

SPECIFICATIONS:

Power
  • 220 VAC, single phase, 50/60 Hz. 
  • Maximum power: 3 kW (20A breaker required)
  • A power cord is included without the plug
Working Temperature
  • 100~1100°C (Continuous)
  • 1200°C (<1 hour) 
Max. Heating Rate
  • ≤ 20 /min


Tube Furnace

 

Temperature Control
  • Built-in PID auto-tune function with overheating & broken thermocouple protection
  • Over-temperature protection and alarm allows for operation without attendant(s)
  • Precise temperature control with less than 1 temperature fluctuation during the dwelling.
  • Optional:  
  • The furnace may be remotely controlled by a PC by installing a communication module in the controller. Please order an MTS-02 control module at an extra cost
  • If you order the PC control module together with the furnace, we will install and test it before shipping. We recommend you to order a software-ready laptop from us 
  •                         
Temperature Uniformity
      
  • 150mm (+/-1°C) heating area with the best temperature uniformity for each heating zone. 
  • A longer area of temperature uniformity (~ 250mm) can be created by setting both zones at the same temperature 
  • Please click the picture below to view the detail of  temperature uniformity profile at 800°C


Flanges & Fittings
(Flange size will be matched with the tube size. Please select in the options bar)



  • Flanges with 1/4” tube fittings and mechanical valves are included on both sides.
  • Flange fixtures and tube covers are installed to provide additional safety protection in case of hydrogen hazards. It is easy to open one side of the flange for quick sample loading
  • Flange with an additional 1/4" thermocouple feedthrough is available at an extra cost for in-situ temperature calibration and measurements.
  • Flange with KF25 port is also available at an extra cost for a quick connection with a vacuum pump.
  • Water-cooled flanges at an extra cost are required if thermal blocks are not inserted. 
              
Gas Delivery
  • The tube furnace comes with a pressure gauge and 0 - 1000 ml/min floating flow meter to precisely control the overall pressure and flow rate in the quartz tube. 
  • Two separated gas delivery lines for hydrogen and inert gases.
  • The inert gas line comes with a mechanical needle valve.
  • The hydrogen line comes with a mechanical needle valve and a solenoid valve controlled by the hydrogen gas detector. 
Hydrogen Detector System
 (included)
  • One Honeywell UL Approved Sensepoint Gas Detector is integrated into the system.
  • Working Temperature Range: 40ºF to +149ºF (-40ºC to +65ºC)
  • Alarm Set Point at 10% (10% LEL). The hydrogen gas delivery line and heating elements will be shut down automatically once the alarm is activated. 
Hydrogen Torch
  • 1/4" O.D. stainless steel tube on the right side is installed as a hydrogen torch with a manual light to safely burn out the hydrogen exhaust. 

Dimension
  • 590 mm(L) x 380 mm(W) x 520 mm(H)

Net Weight
  • ~50 kg
Shipping weight
  • 200 lbs
Warranty
  • One year limited warranty with lifetime support. 
  • The quartz tube as a consumable part is not covered by warranty. 
  • Attention: A free course of safety operation training at the MTI field is required to purchase this furnace. A customer's side training is also available upon request at extra cost. If the users believe that they have sufficient experience and knowledge to handle the furnace in a dangerous situation or emergency, a liability release agreement must be signed at the purchase. MTI Corporation is not responsible for any damages caused by misuse without safety operation training. Any damages caused by the corrosive and acidic gases are not under the coverage of the MTI One Year Limited Warranty.
Compliance

  • CE Certified

 

Warning
  
Click here to learn the installation of a gas regulator
  • Tube furnaces with alumina tubes are designed for use under vacuum < 0.12 atm (absolute pressure) and low  positive pressure 
  • The flow rate of  gases should be < 200 SCCM (or 200 ml/min) for reducing thermal shocks to the tube
  • Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
  • Please note that the regulator for hydrogen gas is different from that for inert gas. 
  •  Quartz tubes may only be safely used up to 1000° under vacuum conditions. 
Manufacturer Part Number: