GSL-KR100 is a combustion gas module to dispose of residual flammable gas via a self-ignition heating coil. It is designed to pair with any MTI tube and atmosphere furnaces. It can also control the gas flow rate and communicate with a Honeywell hydrogen gas detector. It is a good choice for experiments under a reducing environment, such as hydrogen flow. The device also has an RJ45 ethernet port for remote control and monitoring.
SPECIFICATIONS
| Power |
|
| Voltage |
- 110V, 208-240 VAC, Single Phase, 50/60 Hz
|
Features
|
- Applicable to dispose of various residual flammable gas along the exhaust line (designated gas needs to be discussed with MTI in advance).
- Compact, modularized, and compatible with various MTI furnaces.
- Interlocks: in case of emergency, the gas flow and furnace heating will be cut off.
- a flammable gas leak is detected
- Failure of a self-ignition heating coil
- Overpressure
- Out of flammable gas flow.
- One check valve to prevent gas backflow.
- Two solenoid valve allows users to choose the flow type, inert or flammable.
- One mass flow controller, 0-1000 sccm.
-
Double explosion protection: Pressure sensor and mechanical pressure relief valve.
- Constant pressure mode.
- Automated hydrogen treatment.
- Remote control, data monitoring, and log.
- Furnace tube gas purging
|
| Flow rate |
|
| Gas Inlet |
- Two 1/4" compression tube fittings for both inert and flammable gas.
|
Hydrogen Gas Detector
|
The hydrogen gas detector is connected to the control box. The alarm is preset at 20% LEL for hydrogen. If alarmed, the hydrogen gas flow and furnace heating function will be cut off. 
|
PC Control

|
- All operations need to be completed on the host computer
- Manual purging and automatic purging modes are available
- Constant pressure sintering mode: During the heating process, the control system will automatically adjust the air intake and output to keep the heating furnace in a slightly positive pressure state. (Range: 100,000pa-115,000pa)
|
| Over Pressure Protection |
- The electronic pressure transmitter will turn on the exhaust valve to accelerate the exhaust if the set pressure reached
- Mechanical overpressure exhaust valve (0.4 bar relative pressure)
|
| Dimension |
|
| Net Weight |
|
| Safety WARNING |
- If equipped with a tube furnace, a metallic mesh cover must be installed at both ends of the furnace tube in case of emergency
- The furnace and the combustion equipment must be placed in a well-ventilated place, or in a fume hood
- The hydrogen detector must be installed
- Operators must be trained by qualified technicians & follow the safety guideline of the laboratory
|
| Warranty |
-
One-year limited warranty (The warranty does not cover consumable parts such as heating elements, processing tubes, and o-rings, please order the replacement at related products from MTI.)
|
GSL-KR200 is a combustion gas module to dispose of residual flammable gas via a self-ignition heating coil. It is designed to pair with any MTI tube and atmosphere furnaces. It can also control the gas flow rate and communicate with a Honeywell hydrogen gas detector. It is a good choice for experiments under a reducing environment, such as hydrogen flow.