GSL-1100X-S is a 1100oC tube furnace with small foot-print, low weight and low power requirement. It can be easily accommodated in a glove box for material synthesis which vacuum or specific gas conditions.
SPECIFICATIONS:
Furnace Construction
|
|
Chamber Size
|
- Size: 4"x4"x4" or 100x100x100 mm
- Capacity: 65 cubic-inch or 1 liter
-
Click on picture to see the chamber size
|
Power
|
- 110V version: 950 W ( power cord is included )
- 220V version: 1.2KW
|
Voltage
|
AC 110V or AC 220V 50/60 Hz -- please select the voltage in the option bar
|
Max. Temperature
|
1200 °C (< 1 hour)
|
Continuous Working Temperature
|
1100 °C
|
Max. Heating Rate
|
30 °C /min
|
Temperature Controller
|
- PID automatic control and auto-tune function.
- 30 programmable segments for precise control.
- Built in protection for the over-heating and broken thermocouple.
- Large digital control panel (48 x 96 mm) for easy operation.
|
Temperature Accuracy
|
+/- 1 °C
|
Heating Elements
|
Fe-Cr-Al Alloy doped by Mo
|
Thermocouple
|
K type
|
Overall Dimensions
|
- 200 x 230 x 360 mm
-
Click on picture to enlarge and view details
|
Net Weight
|
~ 8Kg
|
Shipping Dimensions
|
17"(L) x14"(W) x 18"(H)
|
Shipping Weight
|
40 lbs
|
Warranty
|
- One year limited warranty with lifetime support
- ATTENTION: Any damages caused by the use of corrosive and acidic gases are not under the coverage of MTI One Year Limited Warranty.
|
Compliance
|
- CE Certified
- All electric components ( >24V) are UL / MET / CSA certified
- The furnace is ready to pass TUV(UL61010) or CSA certification at extra cost. please click marks below to learn details )
|
Application Notes
|
- For preserving the longevity of the furnace components, we recommend the heating rate not to exceed 10 °C /min.
- The cooling rate shall also not exceed 10 °C /min.
- Toxic or explosive gases are not recommended for use with this furnace without necessary safety controls and supervision. Please put furnace into fume hood for better safety ( click picture left )
- Small cracks may appear on the surface of the refractory ceramics over extended use. This is a normal occurrence and the cracks may be repaired with alumina coating.
- May put the small furnace into a standard glovebox via airlock chamber and fire sample under inert gas.
|
Operation Instructions |
|
1100oC