GSL-KR200 is a combustion gas module to dispose of residual flammable gas via a self-ignition heating coil. It is designed to pair with any MTI tube and atmosphere furnaces. It can also control the gas flow rate and communicate with a Honeywell hydrogen gas detector. It is a good choice for experiments under a reducing environment, such as hydrogen flow. 

It is highly recommended to perform inspection on leakage test and purge the whole system before each use, please click this article to see more details about inspection guide.

SPECIFICATIONS

Power
  •  200W
Voltage
  •  110V, 208-240VAC, single phase, 50-60Hz
Features
  • Applicable to dispose of various residual flammable gas along the exhaust line (designated gas needs to be discussed with MTI in advance).
  • Compact, modularized, and compatible with various MTI furnaces. 
  • Interlocks: in case of emergency, the gas flow and furnace heating will be cut off. 
    • a flammable gas leak is detected
    • Failure of a self-ignition heating coil
  • One check valve to prevent gas backflow. 
  • Two solenoid valve allows users to choose the flow type, inert or flammable. 
Flow rate
  • One float flow meter (100-1000 sccm, air)
Gas Inlets
  • Two 1/4" compression tube fittings for both inert and flammable gas.
Hydrogen Gas Detector

  • The hydrogen gas detector is connected externally. If alarmed, the hydrogen gas and furnace heating function will be cut off.
  • The alarm is preset at 20% LEL for hydrogen. 

Dimension

  • 360L × 340W × 400H (mm)

Net Weight

  • About 5Kg

Safety WARNING

  • If equipped with a tube furnace, a metallic mesh cover must be installed at both ends of the furnace tube in case of emergency
  • The furnace and the combustion equipment must be placed in a well-ventilated place, or in a fume hood
  • The hydrogen detector must be installed
  • Operators must be trained by qualified technicians & follow the safety guideline of the laboratory

 

MTI offers a One-Year limited warranty with lifetime support(may cause a fee) for this product.

The MTI One Year Limited Warranty does NOT cover

  • Consumable parts such as Graphite dies, O-rings, crucibles, screws, washers, etc.
  • Any damage and rust resulting from improper storage or maintenance.
  • Any damages caused by the use of corrosive and acidic gases.
  • Any damage resulting from failing to adhere to the individual product's specific operation requirements, such as water cooling, venting, etc.