OTF-1500X-GB is a splitable tube furnace connected to the dual glovebox with the purification system, which allows processing air-sensitive material at the environment of oxygen and humidity less than 1 ppm. The sample after heat treatment can be taken out from the tube furnace to the glovebox for further processing.
SPECIFICATIONS
Dual Glovebox
|
-
Fully openable front window panel (tempered glass, 8 mm thick)
- Dimensions: 2400mm(L) x 800mm(W) x 930mm(H)
- Chamber Material: Stainless Steel 304
- Built-in fluorescent lamp
- Two shelves, 300mm L x 250mm W x 300mm H
- Eight KF40 Feedthrough in the back (can be customized upon request)
- Two pieces of HEPA filters for particle size >0.3 μm
- Large Antechamber Dimensions: 385mm(Dia.) x 588mm(L)
- Small Antechamber Dimensions (for tools): 150mm(Dia.) x 300mm(L)
- The automatic pressure control system is built in to keep the pressure at a constant level.
- A KF50 feedthrough with Manual Gate Valve on the left is standard for use with the furnace.
- For more detailed specifications of the Glovebox, please click this link
|
Purification Unit |
- Water and oxygen concentration: <1 ppm (20°C, 1 atm)
- Working gas: Inert gas such as N2, Ar, He
- Regenerating gas: Mixture of H2 (5%) and 95% working gas
- Note: If you use Ar gas, please make sure the inside equipment is Ar gas compatible.
- Auto regeneration process with user defined parameters.
- Automatically purging function is built-in for quickly reducing O2 < 100ppm & H2O < 100 ppm before cycling.
- Purification Cylinder: made of BASF antioxidant materials and UOP water adsorbent materials.
|
Tube Furnace
|
- Double layer steel casing with air cooling to keep furnace surface temperature lower than 60°C.
- High purity fibrous alumina insulation for Max. Energy saving.
- The special refractory coating on inner to increase furnace life.
- Two pairs of fibrous ceramic tube blocks are included for blocking heat radiation from inside tube. (Ceramic blocks must be fully inserted into the furnace chamber before heating)
- Split cover with Interlock protection, cut the power if door open during furnace heating process.
- 2" Alumina tube with the dimension of 50 O.D x 40 I.D x 1000 L (mm)
- Processing tubes with 3" or 4" are available upon request at extra cost.
- The air-tight flanges with the valve and vacuum gauge are included
- The seperate table is included.
- For more detailed specifications of the furnace, please click this link
|
Vacuum Pump
|
- 240 L/min heavy duty double stage rotary vane pump YTP-550-LD is included.
|
Power |
- Furnace: 208 - 240 VAC, single-phase, 50/60 Hz, 6.0 kW
- Glovebox: 110 VAC or 220 VAC, single-phase, 50/60Hz, 1.7 kW
|
Compliance
|
- Rusting and damage due to improper storage or maintenance are not covered by warranty
- Gloves are consumable items and are not covered under warranty
- NRTL Certification (UL 1450) is available upon request at extra cost
|