VBF-1100X-Q is a 2" ID vacuum tube furnace with the function of fast heating and cooling ( optional water quenching ) for heat treatment of air-sensitive materials. The sample will be placed inside half opened and vacuum-sealed tube, then insert into the pre-heated furnace to achieve fast heating (~ 10oC/S ). After treatment, the tube will be raised up automatically, then insert into an ice tank for quenching ( quickly cooling ). This furnace was designed for LPS powder and can be used for quenching all material under vacuum or controlled atmosphere

SPECIFICATIONS

Power

  • Voltage: Single-phase 110 - 120 VAC / 50/60Hz, (20 A breaker installed)
  • 1500 W max. 
  • Note: Power cable included but without the plug. Please install a plug on your own or click the link below the picture to choose a plug according to your country or your lab requirements


Working Temperature

  • <= 1100°C continuously
  • Max. 1200°C, < 30 minutes
  • Temperature Uniformity: +/- 2°C

Heating Chamber

  • The heating module is made of Ni-Cr-Al resistance wire with a fibrous alumina insulating layer.
  • Quartz tube has a dimension of 50mm O.D. x 44 I.D. x 263 H, mm)
  • SS304 vacuum flange with KF25 Vac. port with valve and gas inlet &outlet valves 
  • Quartz thermal block is included
  • Mechanical vacuum gauge is built-in
  • Vacuum level: 10-6 torr by turbopump and 10-3 by rotary pump and 10-5 torr by a turbopump

Compliance

  • CE Certified
  • NRTL or CSA certification is available upon request at extra cost.

Warning  

Click here to learn the installation of a gas regulator

  • Tube furnaces with quartz tubes are designed for use under vacuum and at low-pressure < 0.2 bars 
  • Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation.
  • A maximum temperature of 1000°C should be imposed on all quartz tube furnaces to avoid the quartz tubes undergoing deformations under vacuum.
  • The flow rate for gasses should be limited to < 200 SCCM for reducing thermal shocks

 

Net Weight:

  • 100 kg

Shipping:

  • Shipping Weight: 250 lbs
  • Shipping Volume: 45"(L) x 28"(W) x 35"(H) 





MTI offers a One-Year limited warranty with lifetime support(may cause a fee) for this product.

The MTI One Year Limited Warranty does NOT cover

  • Consumable parts such as Graphite dies, O-rings, crucibles, screws, washers, etc.
  • Any damage and rust resulting from improper storage or maintenance.
  • Any damages caused by the use of corrosive and acidic gases.
  • Any damage resulting from failing to adhere to the individual product's specific operation requirements, such as water cooling, venting, etc.
Manufacturer Part Number: