VBF-1100X-Q is a 2" ID vacuum tube furnace with the function of fast heating and cooling ( optional water quenching ) for heat treatment of air-sensitive materials. The sample will be placed inside half opened and vacuum-sealed tube, then insert into the pre-heated furnace to achieve fast heating (~ 10oC/S ). After treatment, the tube will be raised up automatically, then insert into an ice tank for quenching ( quickly cooling ). This furnace was designed for LPS powder and can be used for quenching all material under vacuum or controlled atmosphere
SPECIFICATIONS
Power
|
- Voltage: Single-phase 110 - 120 VAC / 50/60Hz, (20 A breaker installed)
- 1500 W max.
- Note: Power cable included but without the plug. Please install a plug on your own or click the link below the picture to choose a plug according to your country or your lab requirements
|
Working Temperature
|
- <= 1100°C continuously
- Max. 1200°C, < 30 minutes
- Temperature Uniformity: +/- 2°C
|
Heating Chamber
|
- The heating module is made of Ni-Cr-Al resistance wire with a fibrous alumina insulating layer.
- Quartz tube has a dimension of 50mm O.D. x 44 I.D. x 263 H, mm)
- SS304 vacuum flange with KF25 Vac. port with valve and gas inlet &outlet valves
- Quartz thermal block is included
- Mechanical vacuum gauge is built-in
- Vacuum level: 10-6 torr by turbopump and 10-3 by rotary pump and 10-5 torr by a turbopump
|
Net Weight
|
100 kg
|
Shipping Weight
|
250 lbs
|
Shipping Volume
|
45"(L) x 28"(W) x 35"(H)
|
Warranty
|
-
One-year limited warranty with lifetime support (consumable parts such as processing tubes, and O-rings are not covered by the warranty, please order replacements at related products below.).
- ATTENTION: Any damages caused by the use of corrosive and acidic gasses are not under the coverage of MTI One Year Limited Warranty.
|
Compliance
|
- CE Certified
- NRTL or CSA certification is available upon request at extra cost.
|
Warning
Click here to learn the installation of a gas regulator
|
- Tube furnaces with quartz tubes are designed for use under vacuum and at low-pressure < 0.2 bars
- Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation.
- A maximum temperature of 1000°C should be imposed on all quartz tube furnaces to avoid the quartz tubes undergoing deformations under vacuum.
-
The flow rate for gasses should be limited to < 200 SCCM for reducing thermal shocks
|
Optional
|
The furnace can upgrade to 4 quartz tubes with rotating mechanism to achieve high throughput sintering
|