Furnace Structure
|
- Double-layer steel casing.
- High purity 1800oC grade fibrous alumina insulation
- Alumina processing tube, vacuum & gas sealing flange with valves, vacuum gauge, and thermal blocks are included for immediate use
- Three graphite electrodes ( 6 mm in diameter) are inserted into the top of the flange for molten salt research under flowing inert gas with a max current of 60A
- The tungsten electrodes are available upon request.
|
Power Consumption
|
Max. 4 KW
|
Input Voltage
|
Single Phase, 220V AC, 50/60 Hz
|
Normal Working Temperature:
|
800-1700 oC
|
Max. Heating / Cooling Rate:
|
5 oC / min above 1200oC and 10 /min below 1200oC
|
Heating Zone
|
- Heating zone length: 200mm
- Constant temperature zone: 100mm within +/-5ºC
|
Heating Elements
|
4 pcs of U-shaped 1900 grade MoSi2
|
Processing Tube
|
|
Temperature Controller
|
- FA-YD518P-AG temperature controller is included.
- Proportional–integral–derivative control (PID control) and auto-tune function
- 30 segments programmed with ramping, cooling, and dwelling steps
- Built-in over-temperature alarm and thermocouple failure alarm
- +/- 1 ºC temperature control accuracy
- Default DB9 PC communication port
- MET Certified
|
Temperature controller (optional)
|
- MTS02-Y Temperature control software kit (for YD518P series controllers) + 15" laptop package is available at the options bar. Click for detailed information about the MTS02-Y software kit.
- You may upgrade the temperature controller to the Eurytherm 3008 Temperature Controller which is fully compatible with LabVIEW. This package includes a Eurytherm 3008 programmer, a communication cable (RS485 - USB), and a software CD. Click the pic below to view the full spec of the Eurytherm 3008 temperature controller.
|
Vacuum Sealing Flanges
|
- One pair of sealing flange with double silicone high-temperature O-ring is included.
- Two SS needle valves and one diaphragm pressure gauge are installed with a flange for immediate use.
- ISO-100K is available.
- Vacuum pressure can reach 50 mTorr by a quality mechanical vacuum pump and 10^-5 torr by a turbopump. Please click the picture below to see the flange assembly details.
|
Water Cooling (Optional)
|
- Water cooling is required for cooling the vacuum flange.
- Please click the picture left to order a recirculating water chiller ( 59L/minute)
|
DC Power Supply (Optional)
|
- A programmable DC power supply is optional up to 30V DC, and 70A Max
|
Vacuum Level
|
- Limited by the vacuum pump
- 50 m-torr can be achieved by the mechanical pump
- 10^-5 torr can be reached by the molecular pump
The vacuum pump is not included, please order separately by clicking the pictures
|
Power Cable & Plug
|
- 10 feet long 10-3 AWG heavy-duty UL-approved power cable is included. Extra length cable can be ordered by clicking picture below-left
-
The plug is not included. Browse our catalog for compatible plugs or obtain one from a local hardware store.
|
Certificates
|
- CE Certified
-
NRTL or CSA certification is available upon request at extra cost. ( please click marks below to learn details )
|
Warning
|
MoSi2 heating element is used in high temperatures under an oxidizing atmosphere. During the oxidizing process, the SiO2 protective film is formed again the heating element from melting during sustained use. The MoSi2 heating element must not be used in the temperature between 400oC and 700oC for a long time, otherwise, the element will be cremated under the strong oxidizing atmosphere at low temperatures.
|